Air Flow Transducer Model 6333
Compact stationary anemometer for remote monitoring of air velocity, temperature and humidity. Model 6333 supports 15 interchangeable probe types, switchable analogue output and RS-485 communication for equipment, facility and IoT monitoring installations.
Model and Probe Overview
| Component | Measurement capability | Range / configuration | Selection guidance |
|---|---|---|---|
| 6333-00 main unit | Stationary transmitter with analogue and digital outputs | 12–24 V DC; switchable 4–20 mA or 0–5 V; RS-485 | Select the probe, cable length, power option and mounting accessories for the installation. |
| Velocity probes | Directional or omnidirectional air-velocity measurement | 0.01–25.0, 30.0 or 50.0 m/s depending on probe | Choose by velocity range, directivity, response time, temperature resistance and probe geometry. |
| Velocity + temperature probes | Simultaneous velocity and temperature measurement | Up to 100°C or 120°C depending on probe | Suitable where process-air temperature must be recorded with velocity. |
| Velocity + temperature + humidity probe | Three-parameter remote monitoring | 0.01–25.0 m/s; 0–60°C; 5–95% RH | Use Model 0975-31 for combined environmental monitoring. |
Key Features
Fifteen compatible probes
Supports directional and omnidirectional probes for velocity-only, velocity-and-temperature, or velocity-temperature-humidity measurement.
Wide velocity coverage
Probe-dependent velocity ranges extend from 0.01 m/s to 25, 30 or 50 m/s for low-flow monitoring through higher-velocity process applications.
High-temperature capability
Selected probes support operating and compensation temperatures up to 100°C or 120°C for demanding equipment and process-air measurements.
Analogue and digital outputs
A DIP switch selects 4–20 mA current output or 0–5 V voltage output, while RS-485 supports digital communication.
Compact stationary design
The approximately 100 × 60 × 25 mm, 200 g main unit can be integrated into equipment or installed using the optional DIN-rail fixing base.
Remote multiparameter monitoring
Suitable probe configurations allow simultaneous output of air velocity, temperature and humidity for facilities, factories, data centres and IoT systems.
Technical Specifications
| Parameter | Unit | Value / Description |
|---|---|---|
| MAIN UNIT — MODEL 6333-00 | ||
| Instrument type | — | Stationary airflow transducer / anemometer for remote monitoring |
| Compatible probes | — | 15 probe types for velocity, temperature and humidity measurements |
| Air-velocity range | m/s | 0.01–25.0, 0.01–30.0 or 0.01–50.0 depending on probe |
| Velocity accuracy | — | Probe dependent; generally ±2% or ±3% of reading, or ±0.02 m/s, whichever is greater |
| Temperature range | °C | 0–60, 0–100 or 0–120 depending on probe |
| Humidity range | % RH | 5–95 with probe 0975-31 |
| Resolution | — | Velocity: 0.01 m/s to 9.99 m/s and 0.1 m/s from 10.0–50.0 m/s; temperature: 0.1°C; humidity: 0.1% RH |
| Display units | — | m/s, °C and % RH |
| OUTPUTS AND POWER | ||
| Analogue output | — | Switchable by DIP switch: 4–20 mA current output, maximum 250 Ω load, or 0–5 V voltage output |
| Digital interface | — | RS-485 |
| Power supply | V DC | 12–24; optional AC adaptor accepts 100–240 V AC input |
| Power consumption | W | Approximately 3 under the manufacturer’s stated reference condition |
| ENVIRONMENT AND PHYSICAL | ||
| Operating environment | — | 0–60°C; 2–95% RH; no condensation |
| Storage environment | — | −20–70°C; 2–95% RH; no condensation |
| Dimensions W × D × H | mm | Approximately 100 × 60 × 25 |
| Weight | g | Approximately 200 |
| Standard accessories | — | Instruction manual and terminal block |
| Optional items | — | Probe, probe cable, AC adaptor, DIN-rail fixing base, probe fixing jig and probe mounting tape |
Velocity accuracy, response time and temperature compensation are probe dependent. Confirm the selected probe model, cable length, output setting, power arrangement and calibration scope before ordering.
Applications
Ventilation and facility monitoring
Provides continuous velocity, temperature or humidity data from selected points in building ventilation systems.
Cooling-air monitoring
Supports remote observation of supply, return and equipment cooling airflow where the chosen probe range and mounting method are suitable.
Process and equipment airflow
Integrates airflow data into production equipment, environmental chambers and process monitoring systems.
Remote data acquisition
Uses analogue outputs or RS-485 communication for controller, data-logger and monitoring-system integration.
Heated-air measurement
Selected probes support velocity and temperature measurement in environments up to 100°C or 120°C.
Three-parameter monitoring
The 0975-31 probe supports simultaneous velocity, temperature and relative-humidity monitoring.